|
MEASUREMENTS
OF DEAD WEIGHT TESTER
PERFORMANCE
USING HIGH RESOLUTION
QUARTZ CRYSTAL PRESSURE TRANSDUCERS
|
|
|
Dr. Richard B. Wearn, Jr. |
Jerome M. Paros |
|
Well Test Instruments, Inc. |
Paroscientific, Inc.
|
|
4520 148thAve. |
4500 148th Ave. N.E.
|
|
Redmond, WA 98052 |
Redmond,
WA 98052
|
|
|
INTRODUCTION |
|
Laboratory users and
manufactures of primary pressure standards are interested in
evaluating dead weight tester performance and making cross
comparisons between different standards.
Quartz crystal pressure transducers have been developed
which may be used to make high speed, high resolution
measurements of dead weight tester performance.
Transducer resolution of parts per billion of full scale
output allows observation of piston-cylinder non-uniformity and
taper equivalent to changes in diameter of one molecular layer.
Pressure fluctuations associated with dead weight tester
rotation, weight bounce, drive mechanism imperfections, and
piston flexing are easily measured.
Errors from these fluctuations can be minimized by
choosing an appropriate averaging time.
These observations have implications for proper design
and use of dead weight tester systems.
|
|
BACKGROUND
|
|
Inherently digital pressure transducers
employing quartz crystal frequency output sensors have been
developed for use in applications requiring high resolution and
precision. Although
these transducers are described in more detail in Reference 1,
the following is a brief summary of the construction, operation,
and performance of the Digiquartz Pressure Instrumentation
manufactured by Paroscientific, Inc. |
| DOUBLE-ENDED TUNING
FORK RESONATOR
|
|
The key sensing element in these
transducers is a vibrating quartz crystal whose resonant
frequency changes with pressure-induced stress.
Figure 1 shows a resonator of the double-ended tuning
fork (DETF) design. It
consists of two identical beams flexing 180 degrees out of phase
to cancel the opposing forces and moments, thus transmitting
very little energy to the mounting pads.
Even though the quartz crystal has a high Q and a
long-lasting resonance, a small amount of energy must be
supplied to achieve and maintain oscillation of the DETF.
Surface electrodes piezoelectrically drive and detect the
resonant frequency through an external oscillator circuit.
The electrode pattern is produced as an integral part of
the photolithographic and chemical milling methods used in DETF
manufacturing (Reference 2).
The resonant frequency of the tines is a function of the
dimensions, composition, and applied load.
As with a violin string, the frequency of oscillation
increases under tension and decreases with compressive loading.
As described in References 3 and 4, the resonators must
be carefully designed to minimize energy losses and to avoid
spurious resonances, which can produce discontinuities in the
frequency versus load relationship. |
|
 |
|
Figure 1.
Double-Ended Fork |
|
TEMPERATURE-SENSITIVE
QUARTZ RESONATOR |
|
The DETF is insensitive to
temperature because the crystallographic orientation and force
producing structure were designed to minimize thermal effects
(Reference 5). The
remaining temperature sensitivity is readily compensated using a
quartz crystal thermometer.
A quartz resonator that is sensitive only to temperature
(not load) is shown in figure 2.
It consists of two torsionally oscillating tines
connected to a mounting pad through a mechanical isolation
system (Reference 6). The
dimensions of the temperature sensitive quartz resonator must be
carefully chosen to avoid spurious modes of oscillation
(Reference 7). These
temperature sensors are manufactured with the same
photolithographic, chemical milling techniques used to make the
DETF resonators. An
oscillator piezoeletrically excites and detects the torsional
tine vibrations. The
change in resonant frequency of the temperature sensitive quartz
resonator can conveniently be used to compensate for thermal
effects in the DETF force sensor and pressure transducer
mechanism.
|
|

|
|
Figure 2. Temperature
Sensitive Quartz Resonator |
|
|
PRESSURE
TRANSDUCER |
|
The pressure-induced load applied to the
force sensitive resonator may be generated by a bellows,
diaphragm, or Bourdon tube. The
higher range designs generally employ a Bourdon tube as the
pressure-to-load converter (Reference 8).
Figure 3 shows a single-turn Bourdon tube restrained by
the DETF resonator. Applied
internal pressure tends to uncoil the Bourdon tube, thus placing
the crystal under tension and increasing the resonators
frequency of oscillation. An
integral torsional tuning fork temperature sensor is used for
thermal compensation. Small,
adjustable masses are positioned such that the center of gravity
of the mechanism coincides with the effective pivot or center of
rotation of the Bourdon mechanism.
Thus inertial forces and torques are reduced to zero and
the transducer has a low sensitivity to linear acceleration and
vibration. A
hermetically sealed case encloses the internal vacuum in which
the resonators operate, thus eliminating air damping and
contamination while ensuring high Q values and a stable
reference for the absolute pressure transducers.

Figure
3. Quartz Crystal Pressure Transducer |
|
INTELLIGENT ELECTRONICS |
|
The quartz crystal transducer can be mated
with a special digital interface board to produce an intelligent
transmitter. Each pressure transducer provides two continuous
frequency output signals, one corresponding to the pressure and
the other to the sensors internal temperature.
The digital board measures the period of these two
signals and calculates fully corrected pressure and temperature.
As shown in Figure 4, the digital board has
a microprocessor-controlled counter and RS232 port.
The microprocessor operating program is stored in
permanent memory (EPROM) and user controllable parameters are
stored in user writable memory (EEPROM).
The user interacts with the transmitter via the two-way
RS232 interface. These
transmitters output fully temperature corrected pressure
information on a two way addressable RS232 bus that can be
interfaced to a computer or stand-alone readout display.
The RS232 interface allows complete remote
configuration and control of all operating parameters of the
intelligent transmitter, including resolution, sample rate,
integration time, and baud rate.
Resolution is programmable from 0.04 to 100 parts per
million depending on system requirements.
Baud rate is user selectable from 150 to 19,200.
Pressure data are available in eight different selectable
standard engineering units.
Up to 15 data readings per second can be obtained with
normal sampling commands. More
than 100 samples per second can be obtained with special burst
sampling commands.
The ability to control the transmitters
resolution and integration time is important in developing a
practical instrument to measure dead weight tester performance. |
|
HIGH RESOLUTION
MEASUREMENTS |
|
The first step in processing the output of
the quartz crystal pressure transducer is to measure its
resonant frequency with a period averaging technique.
The output from the resonator gates a high frequency
clock and the clock pulses are counted.
With an unsophisticated counter-timer scheme, there is
uncertainty of +/- 1 count out of the total number of clock
pulses. The total number of clock pulses equals the clock
frequency multiplied by the integration time ( number of
resonator periods averaged multiplied by the resonator period).
For example, integrating for one second with a 15 MHz
clock yields a frequency resolution of the resonators
output of 0.07 parts per million (ppm).
The quartz crystal pressure transducer is designed to
produce a 10% change in resonator frequency from zero to full
scale applied pressure. Thus
only 10% of the counts are related to pressure and the pressure
resolution would be 0.7 ppm using a 15 MHz clock and update time
of 1 second. Higher
resolution, interpolating start-stop counters are available with
equivalent clock frequency close to the GHz range.
With this improved counting system, the resolution was
found to be limited by a small amount (about 30 nsec) of phase
jitter on the pressure signal waveform caused by cross-talk from
the temperature oscillator.
Although this phase jitter is smaller than the time base
resolution of most counters, the transducer electronics were
modified to provide gating for the temperature oscillator.
With the temperature oscillator gated off while measuring
the pressure signal, resolution was improved by more than a
factor of 4 for any integration time and was least count limited
by the interpolating start-stop counter.
Actual sensor performance may be even better than listed
in Table 1.
|
Integration Time (sec) |
Resolution |
| 0.1 |
2.5x10-7 |
| 1 |
2.5x10-8 |
| 10 |
2.5x10-9 |
Table 1. Pressure Resolution Temperature Resolution Gated Off
Tests were then run to show that the
transducers can measure real pressure changes that exceed the
noise levels shown in Table 1.
Figure 5 shows a 15,000 psia (103 MPa) transducer
tracking small weight wobble pressure fluctuations on a DH
Instruments primary dead weight pressure standard.
Sensor output was integrated for 0.1 sec.
Pressure fluctuations of a few parts in 107
are readily resolved.

An even more remarkable set of measurements
is shown in Figure 6 in which small atmospheric pressure changes
over a five minute interval are tracked simultaneously with a
barometer and with a 10,000 psi (69 MPa) transducer.
Real pressure changes of the order of 0.00005 psi (5
parts per billion full scale) are being tracked with a 10,000
psi transducer! Counter
integration time was 10 seconds.

|
| PERFORMANCE
MEASUREMENTS OF PRIMARY DEAD WEIGHT PRESSURE STANDARDS |
|
To demonstrate the high resolution
capabilities of the quartz crystal pressure transducers, they
were used to study the performance of the DH Instruments primary
dead weight pressure standard.
Figure 7 shows the piston area variation
with float height for three different pistons, as determined
from pressure measurements as a function of height.
About 15 minutes were required for three profiles on a
single piston. A
few parts in 107 changes in piston area with height
are easily resolved, corresponding to an equivalent change in
diameter of one molecular layer.
The 100 psi/kg piston is remarkably uniform in area along
its length, varying less than 1 ppm in the central 0.5 cm.
Not surprisingly, the smaller diameter, higher pressure
pistons tend to show greater variation in area.
The data suggests some interesting applications for
screening pistons and monitoring wear patterns with time.
For example, is the 5 ppm relative dip in the central
region of the 20,000 psi piston caused by piston wear?
This is a steel piston in a carbide cylinder and so would
wear to a smaller effective area.
Although these are all very high quality pistons, with
variations less than one percent of a wavelength of light, their
minute departures from perfection are readily measured with the
quartz crystal pressure transducers.
Less expensive pistons from other manufacturers
frequently show changes of 200 ppm in area along the piston.

|
|
CROSS COMPARISONS OF
PISTON AREAS |
|
DH Instruments provided three well used
pistons of different ranges.
Normally, one compares piston areas by cross-floating
using two dead weight standards.
By switching a transducer back and forth between two dead
weight testers using valves, one would be able quickly to
compare two pistons to one part per million; however, as only
one dead weight standard was available, a much more difficult
procedure was attempted. Pressure
was applied at 8,500 psia with the first piston and measured
with two of the quartz crystal pressure transducers as transfer
standards. The
system was depressurized, the piston was changed, and the
measurements were repeated with the remaining two pistons in
turn. The total
measured spread among the three pistons (after allowing for
differences in nominal piston area using the manufacturers
data) was 12.9 ppm, even though two of the pistons had not been
calibrated in two or three years.
This represents the total cumulative error from the
manufacturers original area measurement, wear with time, mass
errors, and transfer sensor errors. |
|
DEAD WEIGHT TESTER
PRESSURE FLUCTUATIONS |
|
The pressure output of primary dead weight
pressure standards is not precisely constant in time.
High quality standards typically have short period
pressure fluctuations of 10 ppm.
Less accurate standards may have fluctuations of 100 ppm
or more.
Figures 8, 9, 10, and 11 show fluctuations
observed for various pistons at pressure from 6000 psia to
35,000 psia. The
dead weight standard used is a DH Instruments Model 5306, which
is representative of a premium class standard.
Output pressure was measured with Paroscientific Inc,
Model 15K quartz crystal pressure transducers; read out with a HP 5384A counter, with an integration time of 0.1 seconds.
For optimum resolution, data were taken with the sensor
temperature oscillator gated off.
Each run consisted of 100 consecutive period measurements
taken at a rate of approximately 4 points per second.
Figure 8. DWT Fluctuations at
6,000 PSIA Piston S/N 1339
Figure 9. DWT Fluctuations at
10,000 PSIA Piston S/N 1064

Figure 10. DWT Fluctuations at
30,000 PSIA Piston S/N 1340

Figure 11. DWT Fluctuations at
35,000 PSIA Piston S/N 1340
The observed fluctuations often occur at
periods associated with weight rotation, weight bounce, or
pendulum oscillations of the weights flexing the piston.
Note that the Figures 10 and 11 show a pronounced
response at the weight rotation frequency and its second
harmonic, but that other runs show more varied frequency
content. Some show
pronounced changes in amplitude with time.
Although more data are needed, indications
are that the dead weight standard short period fluctuations
result primarily from the weight drive and weight suspension
mechanisms. Now
that sensors are available that can measure these fluctuations,
it should be possible to develop primary pressure standards with
improved performance. |
|
|
CONCLUSION |
|
Current premium class primary dead weight
testers have variations in output pressure with float height and
time, which are readily measured with quartz crystal pressure
transducers. With
these new sensors as a testing tool, it should be possible to
develop primary pressure standards with improved performance.
The sensors also make it easier to cross-check piston
sets against a master reference piston and to screen them for
nonuniformity. |
|
|
REFERENCES |
-
Busse,
D.W., Quartz Transducers for Precision Under
Pressure, Mechanical Engineering, Vol. 109, No. 5, May
1987.
-
EerNisse, E.P., Minature Quartz
Resonator Force Transducer, U.S. Patent
4,215,570 Aug. 5, 1980.
-
Paros, J.M. and Busse, D.W.,
Longitudinal Isolation System for Flexurally Vibrating Force
Transducers U.S. Patent 4,321,173 Mar. 23, 1982.
-
EerNisse, E.P and Paros, J.M., Resonator
Force Transducer U.S. Patent 4,372,173 Feb. 8,1983.
-
Paros, J.M., Isolating And Temperature
Compensating System For Resonators, U.S. Patent 4,406,966
Sept. 27, 1983.
-
Paros, J.M., Wearn, R.B., and Tonn, J.F.,
Mounting and Isolating System for Tuning Fork Temperature
Sensor, U.S. Patent 4,706,259 Nov. 10, 1987.
-
EerNisse, E.P. and Wiggens, R.B.,
Resonator Temperature Transducer, U.S. Patent 4,593,663
June 3, 1986
-
Paros, J.M., Digital Pressure
Transducer, U.S. Patent
4,455,874 June 26, 1984.
|
|
©2007
Paroscientific, Inc.
|
|